2
If flammable substances are present before you apply system power, do not
plug in or turn on the system.
If you detect flammable substances after the system has been turned on, do
not touch any of the controls, switches, knobs or cables. Evacuate all personnel
immediately, then ventilate the room to clear the air of the flammable vapor or
gas. Remove any volatile liquids that are producing flammable vapors to a safe
storage area.
Electrical Safety
Electrical circuits inside the equipment use voltages that are capable of causing
serious injury or death from electrical shock.
WARNINGS: This equipment contains high power electrical components and only
personnel familiar with the circuitry and its operation should provide
service. Some components may maintain a charge with the power cord
disconnected.
Observe safety precautions when measuring high voltage electronics.
Steps within this procedure produce x-rays. Use appropriate precautions.
Areas where hazardous voltages exist include:
Transformers - 100 - 240 volts
Video Monitors - 100 - 240 volts and High AC/DC Voltages
Image Intensifier - 50,000 volts
Generator - 110,00 volts
•
Observe the following precautions when you remove system covers:
•
Observe the two person rule when working near lethal voltages. A person
who is familiar with the emergency power removal procedure must be in
attendance. This person must remain clear of the machine and be prepared
to turn it off in an emergency.
•
Be familiar with the components being serviced and the locations of
hazardous voltages in the assembly where you work.
•
High voltage cables and capacitors can retain a charge even when you
remove all power from the system. Avoid touching these unless you are
certain the charge has dissipated to ground.
•
Some high energy capacitors in the system should be shorted to ground
through an adequate resistance to avoid a burn hazard.
•
A protective earth ground stud, located on the bottom left side of the
system base, and marked with the international symbol shown below on the
left. It also has a distinctly colored decal at the base of the stud as shown
below on the right.
e l
Grn
PROTECTIVE EARTH SMBOL
PROTECTIVE EARTH STUD
Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800