13. Screw the t-handle into the piston on the outlet side.
14. Manually push the piston to the bottom of the cell. This will force any
remaining hydraulic fluid back into the pump.
15. Close the pressure relief valve on the pump and close the inlet ball valve.
16. Disconnect the pump from the inlet valve assembly.
17. Remove the inlet valve assembly, cell cap, and piston. Be aware that
some hydraulic fluid will remain in the cell.
18. Clean and dry the entire cell assembly. Pay close attention to the threads
on the cell body and cell caps. Also make sure to clean the outlet valve
assembly. Inspect all o-rings and replace any that show signs of damage
or wear.