Error messages will be shown on the top line of the display and will overwrite the DIO status or time and date
information if they have been enabled in setup. After the message is cleared, the display reverts to the previous
data shown before the error was detected. An example of an error message display is given in
Figure 5-1 and the possible error messages are listed in Table 5-3 with a brief description of the error.
Figure 5-1: Error Message Display
Table 5-3: Error Codes
Error Message
Description
Input string illegal
An alpha-numeric value was entered where only a
numeric value is valid
Preset tare failed
Command failed - .
exceeds capacity limit
Unit switching would result in a value too large to
display
exceeds zero range (+)
ZERO key pressed and weight is above capture
range
exceeds zero range (-)
ZERO key pressed and weight is below capture
range
illegal scale mode
ZERO key pressed while terminal is in net mode
illegal tare value
invalid increment value
invalid tare function
invalid zero function
keypad tare disabled
A preset tare was attempted while the keypad tare
function is disabled
motion
Scale motion inhibited completion of command
no demand output
A print was requested without a demand connection
power-up zero not captured
scale in x10 mode
Command not possible while scale is expanded x10
scale not ready
Print requested and print interlock or auto print
reset requirements have not been met
scale over capacity
Print was requested while scale was over capacity
scale under zero
Print was requested while scale was blanked under
capacity
tare exceeds limit
Preset tare value is greater than scale capacity
tare not enabled
TARE
key pressed while tare is disabled
tare too small
TARE
key was pressed while weight was less than
1 display increment
tare under zero
TARE
key was pressed while weight was below
zero
Coin cell voltage low
Contact service
5.3.5
Internal Diagnostics
The T72XW terminal provides several internal diagnostic tools that are accessible in setup mode.These tests are
intended to assist in diagnosing whether a problem is internal or external to the terminal.
To access these tools, enter the Operator menu then the
SETUP
mode and navigate to the Maintenance menu.
Open the Maintenance block then the Diagnostic block.