EN-10
a-AP30
5. TECHNICAL DATA
5.1 Specification
Equipment Ratings:
Indoor use only
Altitude:
2,000m/6,562ft
Operating temperature: 5
℃
to 40
℃
(41
℉
to 104
℉
)
Humidity:
Maximum relative humidity 80% for temperatures
up to 31
℃
(87.8
℉
) decreasing linearly to 50%
relative humidity at 40
℃
(104
℉
).
Electrical supply:
4 x AAA (LR03) Batteries.
Pollution degree:
2
Table 5-1 Models Specification
Model
AP30TURL
AP30TURH
Lamp
LED IR
LED IR
Measure Range
0 ~ 200 NTU
0~1000 NTU
Calibration
Up to 4 Points
Max 8 Point Calibration
Accuracy
≤0.2NTU
(
0~5NTU
)
or ±5%
(
5~200NTU
)
≤0.3NTU
or ±8% F.S.
Repeatability
≤0.1NTU
,
0~5NTU
;
Reading ±1%
,
5~200NTU
≤0.1NTU
,
0~10NTU
;
±1%
,
10~200/1000NTU
Zero Drift
±0.1 NTU /30 min
≤0.3% F.S /3
0 min
Resolution
0.01 (0~9.99 NTU),
0.1 (10~99.9 NTU),
1 (100~200 NTU)
0.01 (0~9.99 NTU),
0.1 (10~99.9 NTU),
1 (100~1000 NTU)
Operating Temp
5 ~ 40
℃
Storage Temp
-10 ~ 55
℃
Humidity
0 ~ 80 % R.H. (non-condensing)
Working Altitude
Up to 2000m
Power
4 (AAA)
Batteries (Included)
Dimensions
170 × 72 × 44 mm
Weight
290 g