EN-33
Adventurer Pro
GLP DATA
User ID
Set...
Proj ID
Set...
Time
Type-
12hr
, 24hr,
Set..., Adj-60, +60
Date
Type
mdy
,...dym
Set...
End GLP Data
User ID
Used to enter a User ID up to 10 characters.
Project ID
Used to enter a Project ID up to 10 characters.
3.7.8 RS232-1 and RS232-2 (Cont.)
When the balance is operated from batteries,
the menu item is available and the default setting
is Off. To enable COM1, Power must be turned
On.
Baud
Baud rates of 600, 1200, 2400, 4800, 9600
and 19,200 are available.
Parity
Parity settings of 7 even, 7 Odd, 7 No Parity
and 8 No Parity are available.
Handshake
Settings of Off, XONXOFF and (for RS232-1
only) Hardware are available.
End RS1 (RS2)
Pressing the
Yes
button will advance to the
RS232-2 (or GLP Data) submenu. Pressing
No
returns to the Power (or Baud) menu item.
3.7.9
GLP Data
Entering a User ID or Project ID
Enter the user or Project ID one character at a
time. The character to be entered is highlighted
by a blinking cursor (underscore). Press the
No
button repeatedly to scroll through the list of
available characters: (space), -, 0 through 9, A
through Z. Press the
Yes
button to select the
displayed character and move the cursor one
position to the right. After the tenth character is
entered, the ten character ID will blink. Press
the
No
button to change the displayed ID or the
Yes button to accept the ID and proceed to the
next menu item.
Time
Type
Set the type to 12 hours or 24 hours.
Set
Set the current time using the format selected
in the type menu item.
Adjust
Enter an adjustment value from -60 to + 60
seconds per day.
Date
Set the date type: M/D/Y, D/M/Y, Y/M/D,
M/Y/D, Y/D/M, D/Y/M and the actual date.
End GLP Data
Press the
Yes
button to advance to the GLP
Print submenu. Press the
No
button to return to
User ID.