BW SERIES COMPACT BENCH SCALE EN-3
2.
INSTALLATION
2.1 Unpacking
Unpack and verify that the following components have been included:
• BW Scale Unit
• Weighing Pan
• AC-DC Adapter
• Instruction Manual
Save the packaging material. This packaging ensures the ideal protection for the
storage or transport of the product.
2.2 Transport Protection
BEFORE USING THE SCALE:
Remove the 3 shipping protection screws on
the spider sub-platform, then remove the 3 plastic
column inserts under the spider. Re-install the
protective components if shipping the scale to
ensure proper transport protection.
2.3 Selecting the Location
Operate the scale on a firm, level surface. Avoid locations with rapid temperature
changes, excessive air currents, vibrations, electromagnetic fields, heat or direct
sunlight.
2.4 Leveling the Scale
Adjust the leveling feet* until the bubble is centered in the circle of the level indicator
(located on the front panel). Ensure that the scale is level each time its location is
changed.
*
IMPORTANT
: To allow proper air flow and minimize condensation inside
the scale, unscrew the 2 rear leveling feet slightly, then level the scale accordingly
for normal operation. However, when cleaning the scale, all 4 leveling feet must be
screwed in tightly to avoid entry of water inside the scale.