44
CD Indicators
5.2 Error Codes List
Error Codes List
The following list describes the various error codes and which can appear on the display and the suggested remedy.
Tare Errors
2.0
Tare timeout - Indicator unstable during tare attempt.
2.1
Preset tare timeout - (more than 4 seconds between key presses).
2.2
Zero timeout - Indicator unstable during zero attempt.
Calibration Errors
3.0
Improper or missing mass used for calibration. Recalibrate with correct masses.
User Errors
4.4
RS232 buffer full.
5.1
Scale settings below minimum resolution. (Full Scale/Graduations < 500).
5.2
Too many divisions allowed w/current Grad and Full Scale settings. Increase Grad size, reduce capacity, or
recalibrate.
5.3
Unit switching error - cannot display weight in selected unit based on current settings for Cal Unit or grad.
See chart for legal display units.
7.0
User entry out of bounds. -- value entered for Hour, Minute, Month, or Day is out of range.
7.1
Factor entered for Custom Units <0.1 or > 1.999999.
7.2
Number outside of display capacity.
7.3
Value entered for preset tare is > Full Scale, or is not a multiple of the display increment (see chart below).
7.4
Value entered for Full Scale is > 1999999 or =0.
7.5
Value entered for User Calibration or Calibration Test Weight is > Full Scale or < 10% of Full Scale.
7.6
Value entered for Print Interval is < 1 or > 3600 seconds.
7.7
Value entered for Check Weigh point is > Full Scale, = 0 or is not a multiple of the display increment (Grad).
7.8
Value entered for minimum Check Weigh point > maximum Check Weigh point.
Over-Under Load Errors
8.2
Power-on zero out of range. Weight on scale base greater > zero (2%, 18%, 100%) during power up).
8.3
Capacity exceeded. (weight on pan > Full Scale + display increments).
8.4
Underload condition on scale base (weight on pan < -5 display increments).
CheckSum Errors
9.1
Bad factory checksum. If error persists, have the Indicator serviced.
9.8
User calibration data failed checksum.