Defender 2200 Bench Scale
EN-15
3.4 Setup Menu
When the Indicator is used for the first time, enter this menu to set the
Capacity and Graduation.
Reset
Legal for Trade
Calibration Unit
Capacity
Graduation
Power On Unit
Zero Range
Retain Zero Data
End Setup
No
, Yes
Off
, On
kg¸
lb
5
…20000
0.0005
…20
Auto
, kg, lb, g, oz
2%
, 100%
Off
, On
Exit SETUP menu
3.4.1 Reset
Reset the Setup menu to the factory defaults.
No = not reset.
Yes = reset.
3.4.2 Legal for Trade
Set the legal for trade status.
OFF = off
ON = on
Turning on the “LFT” menu setting has the following effects:
• Zero-range is set and locked on “2”.
• Auto Zero Tracking is set and locked on 0.5d
• The lb:oz unit is not available as a power-on setting.
3.4.3
Calibration Unit
Set the unit during calibration.
CAL UN kg = Calibrate using kg weights
CAL UN lb = Calibrate using pound weights
3.4.4
Capacity
Set the scale capacity from 5 to 20000. Refer to the Setup Table 3.3 for available settings.