EN-20 Defender 5000 Indicators
3.3.4 GEO Adjustment
Set the GEO factor that corresponds to your location.
GEO codes are numbered 0-31.
Long press the button
to enter the Main
Menu. Select the menu item Calibration by
pressing the Softkey corresponding to the icon
.
Scroll to GEO Adjustment using the Softkey
corresponding to the icon
.
Press the Softkey corresponding to the icon
to edit the GEO value. Press the button
and enter the desired value using the
alphanumeric keypad. After editing, press the
Softkey corresponding to the icon
to exit
the menu.
Note:
See table 9-4 for GEO values.
Put the calibration mass on the pan, and then press
the Softkey corresponding to the icon
for
confirmation. To change to a different calibration
point, input the value desired, and then place the
corresponding weight on the pan for calibration.
The message 'Completed' will be displayed on the
screen.
Exit Linearity Calibration by pressing the Softkey
corresponding to the icon
.
To return to the Main Menu, press the Softkey
corresponding to the icon
.