Defender 5000 Indicators
EN-51
Definitions:
Weight
- Up to 11 characters, right justified, - at immediate left of most significant character (if
negative).
Unit
- Up to 5 characters, right justified. If the Unit in the Print Content menu was set to OFF, the
unit will be removed in the weight string and replaced by spaces.
Stability
-
“?” character is printed if not stable. If weight is a space is printed.
T/N/G/PT
-
“T” is printed for a tare weight, “N” printed if weight is net weight, ‘G’ or nothing printed if
weight is a gross weight, 'PT' is printed if the tare weight is Pre-set Tare.
Application Status (for Check)
– Fixed to 6 characters. Display status like " Under", "Accept" and
"Over" for check weighing.
Terminating Character(s)
- terminating character(s) printed depending on FEED menu setting.
Printout string for the lb:oz unit
Field
Weight1 Space Unit1 Space Weight2 Space Unit2 Space Stability Space G/N Space Message Term.Char(s)
Length 4
1
2
1
7
1
2
1
1
1
1
1
5
2
The printout string has a fixed length of 28 characters.
Each Space field is a delimiting space used to separate the other fields.
The Weight1 field is 4 right justified characters. If the value is negative, the '
‐
' character is
located at the immediate left of the most significant digit.
The Unit1 field is 2 left justified characters.
The Weight2 field is 7 right justified characters.
The Unit2 field is 2 left justified characters.
The Stability field is 1 character. A space is printed if the weight value is stable. A '?' is printed if
the weight value is not stable.
The G/N field is 1 character. 'G' is printed for a gross weight. 'N' is printed for a net weight.
The Message field is 5 left justified characters.
Note:
The Termination Characters Carriage Return and Line Feed are appended to the printout.
5.6 Printout Examples
Setup in Menu
Print out
{String 1} {New Line}
{String 2} {New Line}
{String 3} {New Line}
{New Line}
{Time} {3 spaces} {3 spaces} {Date} {New Line}
{ID} {New Line}
{Result} {New Line}
{New Line}
{String 4} {New Line}
{String 5} {New Line}
{End of template}
OHAUS CORPORATION
7 Campus Drive
Suite 310
10:01 04/22/2016
50
500.0 g
Signature__________
Verified by_________