Defender 5000 Indicators
EN-41
4.3.3 Application Setup
The Application can be customized for various user preferences.
Press the Softkey corresponding to the icon
to enter
Configuration Setup
.
The
Configuration Menu
is now displayed.
Select the list item and press the Softkey
corresponding to the icon
,
to change the
setting as desired.
To return to the Application home screen, press
the button corresponding to the icon
.
The Check Configurations are defined below
(defaults in
Bold
).
Configure Item
Option (
Bold
is default)
Description
Check Mode
Weighting
/Counting
Weighing
: Check weighing mode.
Counting: Check counting mode.
Auto Tare
Off
/On
Off
: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a
container automatically.
Accumulate
Off
/Manual/Automatic
Off
: The Softkey
"∑" does not display.
Manual: The Softkey
"∑" displays. User can press the key
to accumulate the stable weight.
Automatic: The Softkey
"∑" displays. The weight will be
accumulated automatically.
Note:
The load to be accumulated has to be >= 5d and the next
accumulation can only start once the pan has been cleared.
When LFT is ON(
no such limit when LFT is OFF or the
approve mod is OIML)
,
a. gross weight and net weight cannot be accumulated;
b. gross 0 has to be reached before a new sample
accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated;
b. a new sample can be accumulated after reaching gross
0 or net 0.
Audible Signal
Off
/
Under and
Over/
Accept
/Under/Over
Off: No beep.
Under and Over/
Accept
/Under/Over: Beep when
reaching the selected check point.
Auto Opt.
Off/
On
Off: Auto Opt. is off.
On
: The APW will be optimized automatically when doing
count weighing.
Opt. Beep
Off/
On
Off: Opt. Beep is off.
On
: When the APW is optimized, the beeper will beep
once.
APW Auto Save
Off/
On
Off: APW Auto Save is off.
On
: If the APW is get from sampling, and a counting
library is selected, the new APW will save to library after
optimized.
Note: It will be hidden when "Auto Opt." is off.