CHAPTER 4 TESTING
4.5.3 Full Load Test
1. After the Off Center Load Test has been passed, test the Base’s full capacity
according to the Specification Tables 1-1 through 1-6. The Base must meet all
specifications as listed.
2. If the Base fails the Full Load Test, check and set the Overload Stops.
(See Section 3.3.)
Ohaus Corporation www.ohaus.com
4-6
Defender™ H-Q-W Series Service Manual