3-2
CHAPTER 3 SCALE TESTING AND CALIBRATION
3.3 TESTING THE LOAD CELL ASSEMBLY
This section contains two methods of testing the Load cell.
3.3.1 Resistance Test
1. Refer to Section 4.1 and remove the Top Housing.
2. Refer to Figure 3-2 and record color code and position of the wires from the Load Cell.
3. Unsolder the Load Cell cable from the Main PC Board. Using an Ohmmeter, measure in accordance
with Table 3-2 Load Cell Resistance Readings on the cable. If the resistance readings vary more
than the table readings allow, replace the Load Cell in accordance with the procedure in Section
4.4 or 4.5. This resistance test is useful in determining if the Load Cell has been severely damaged
by a short or open circuit which has occurred on the cell or wiring. If the Load Cell resistance
readings are good, continue with this procedure.
TABLE 3-2 LOAD CELL RESISTANCE READINGS
MEASUREMENT POINTS
READING
Wires 2 & 5
415 + 15 Ohms
Wires 3 & 4
350 + 3 Ohms
NOTE
: Wires 2, 3, 4 and 5 must be isolated from chassis ground and shield. Resistance should be
greater than 5 megohms. A lower reading indicates a short in either in the Load Cell cable or the Load
Cell.
4. Reconnect Load cell cable to the Main PCB.
3.3.2 Excitation and Output Voltage Test
NOTES
: 1. The following steps involve power applied to the scale.
2. Load Cell solder contacts can be used as measuring points. See Figure 3-2.
1. Insert the Pan Support into the Load Cell Frame and place the Pan on top.
2. Turn on power to the scale.
3. With no load on the Pan, first measure the excitation voltage across point 2 and 5 Load Cell
connection on the PCB. This voltage must be 5.0 Volts dc with the Load cell connected. If the
voltage is lower, disconnect the Load Cell cable from the PC Board and measure again. If the
voltage is 5 Volts dc, the Load Cell is defective and must be replaced. If the voltage remains low,
the PC Board is defective and the Main PC Board must be replaced. For Main PC Board
replacement, see Section 4.3.