4-14
CHAPTER 4 REPAIR PROCEDURES
3. Remove the Battery from the Bottom Housing.
4. Install a replacement Battery into the same position as the original Battery. The positive terminal
(Red) should be facing up.
5. Connect the wires to the Battery. Make sure the Red wire is attached to the positive terminal on the
Battery and the Black wire is attached to the negative terminal on the Battery.
Note
: The Battery is held in place by the Top Housing that has a cutout and foam backing. When the
Top Housing is installed, the Battery is held in place.
CAUTION: Dispose of the Battery according to local regulations for hazardous materials.
Do not incinerate, crush or throw out in regular garbage containers. Battery contains an
acid solution.
6. Reassemble the scale and connect the ac adapter to the scale and plug into a power source. Allow
the battery to charge for at least 12 hours with the scale off. A new battery requires a long charge
period to obtain maximum service when operating the scale for the first time.
4.6
REPLACING THE BATTERY (Cont.)
2. Disconnect the two wires from the battery. See Figure 4-19 which is a close-up view of the
battery.
Figure 4-19. Battery Connections.