EN-108
EXPLORER
BALANCES
5.14.8 Lock Communication
Set the status.
OFF
= the Communication menu is not locked.
ON
= the Communication menu is locked.
5.14.9 Lock Library
Set the status.
OFF
= the Library menu is not locked.
ON
= the Library menu is locked.
5.14.10 Lock I/O Settings
Set the status.
OFF
= the I/O Settings menu is not locked.
ON
= the I/O Settings menu is locked.
5.14.11 Factory Reset
Set the status.
OFF
= the Factory Reset menu is not locked.
ON
= the Factory Reset menu is locked.
5.15 Securing the Menu and Key Lock menu settings
A slide switch is used to secure the Lock menu settings. When the switch is set to the ON position, the Lock menu
settings may be viewed but not changed. This switch is located under the Base.
Set the position of the switch to ON by sliding the external Lock Switch to LOCKED as shown in Section 6.
When the switch is in the ON position, the start up display includes the LOCK ON message.
Note:
This switch is also used in conjunction with the Legal for Trade menu item. When the Legal for Trade menu
is set to ON, the switch must be set to the ON position to prevent calibration and changes to metrologically
significant settings. Refer to Section 6 for more information.