EN-66
EXPLORER
BALANCES
A numeric input window is displayed.
Key in the desired value, then press
9
.
The display returns to the previous screen with the new value
highlighted.
To return to the Density
Determination
home screen,
touch
Return to Application
.
Press
Begin Density Determination
to start the process.
Step 1 of 2 – Weigh the Sinker in Air.
Follow the screen instructions, then press
Accept
to store the sinker weight (“in air”).
Step 2 of 2 – Weigh the Sinker Submerged in the Test
Liquid.
Follow the screen instructions, then press Accept
to store the sinker weight (submerged in liquid).
Once the necessary weights have been determined,
the density of the Liquid sample is displayed in
g/cc
(along with the weight in air, weight in water)
on the Application screen.
Press
Begin Density Determination
to reset all the weight
values and restart the process.