EN-78
EXPLORER
BALANCES
To adjust the water temperature, touch the
Water Temp
button.
The balance calculates water density based on the water
temperature value entered. Measure the actual water
temperature using a precision thermometer.
A numeric input window is displayed.
Key in the desired temperature, then press
Done
.
The display returns to the previous screen with the new value
highlighted.
Continue to enter the required information
following the above procedure:
Pipette Inaccuracy
, via the
Inaccuracy
button
Barometric Pressure
, via the
Barometric Pressure
button
Pipette Imprecision
, via the
Imprecision
button
Pipette Name
, via the
Pipette Name
button
Pipette Number
, via the
Pipette number
button
When finished, press
Return to Application
.
The Application home screen appears, and you can begin the
Pipette Adjustment process.