EXPLORER BALANCES
EN-89
5.2 Main
Menu
The Explorer balance Main menu structure is illustrated below.
Calibration User
Settings
Balance
Setup
Application
Modes
Weighing
Units
GLP and
GMP Data
Communication
Library
I/O
Settings
Diagnostics
Factory
Reset
Lockout Information
5.3 Calibration
Explorer Balances offer a choice of six calibration methods: Internal Calibration, AutoCal Calibration,
Calibration Adjust, Span Calibration, linearity Calibration, and User Calibration.
Note:
Do not disturb the balance during any calibration.
5.3.1 Calibration
sub-menu
AutoCal™
Internal
Calibration
Automatic
Calibration
AutoCal™
Adjustment
Span
Calibration
User
Calibration
Calibration
Test
5.3.2 Internal Calibration (AutoCal™)
Calibration is accomplished with the internal calibration mass. Internal calibration can be performed at
any time, provided the balance has warmed up to operating temperature and is level.
With the Balance turned ON and no load on the pan, touch
AutoCal
™
Internal Calibration
.
The Balance begins to calibrate.
Alternately, touch the shortcut
Calibration
Icon from any Application home screen
to initiate an Internal Calibration.
The display shows the status, then returns to the current application.
To cancel at any time, press
Cancel
.
5.3.3 Automatic
Calibration
When
Automatic Calibration
is set ON, the balance performs a self calibration when a measured
predefined change occurs (in temperature or time). AutoCal will automatically calibrate the Balance (using
the internal mass) each time a change in temperature significant enough to affect accuracy, or after every
11 hours.