EN-108
EXPLORER BALANCES
5.5.6
Functions Configuration
Touch
Functions Configuration
to see a table of Selected and
Available functions. Press
>
to move a function from
Available
to
Selected
, and press
<
to move them back.
To move all functions to the right, Press
>>
and press
<<
to move them back.
5.5.7
TouchLess Sensor Setup
If desired, assign
,
Zero
,
Tare
,
Calibrate, Automatic
draftshield door
or
Draft Shield Light
to the Touchless Sensors
Terminal Right
,
Terminal Left
,
Base Right
, and
Base Left
.
Default setting for Automatic draftshield door models:
Terminal left: left door
Terminal right: right door
Note:
EX12001, EX24001 and EX35001 do not have Base
sensors.
5.5.8
Draft Shield light
Set the status of the lighted Draft Shield feature.
OFF = disabled.
ON
= enabled.
5.5.9
Ionizer (seconds)
Set the working time of the ionizer, 3 or 5 seconds.
Normally, 3 seconds is enough to remove the static electricity from
a sample.
However, if the air is very dry, or if the sample volume is large, a
longer working time might be used: 5 seconds.
Note:
In order to get the best effect, it's recommended to position
the samples at the center of the glass shelf.