EN-110
EXPLORER BALANCES
5.6.6
Auto Tare
Set the automatic Tare.
OFF
= disabled.
ON = enabled.
5.6.7
Gross Indicator
Set the symbol displayed for gross weights.
OFF
= no symbol is displayed.
GROSS = the G symbol is displayed.
BRUTTO = the B symbol is displayed.
5.6.8
Legal For Trade
Use this menu to set the Legal for Trade status.
OFF
= standard operation.
ON = operation complies with Weights and Measures regulations.
Note:
When Legal for Trade is set to ON, the menu settings are affected as follows:
Calibration Menu:
Automatic Calibration is forced to ON and hidden. AutoCa
l
TM
internal Calibration,
and Calibration Test functions are available. All other functions are hidden.
For EX…N… models:
-
Automatic Calibration will be locked at its current setting.
-
If you set AutoCal
TM
internal Calibration to be ON before you turn on Approved
Mode, AutoCal
TM
internal Calibration menu will still be available.
-
If you set AutoCal
TM
internal Calibration to be OFF before you turn on Approved
Mode, AutoCal
TM
internal Calibration menu will be locked.
Balance Setup Menu:
Filter Level is locked at the current setting.
Stability Indicator Range is locked at 1 Division.
Auto Zero Tracking is limited to 0.5 Division and OFF.
Auto Tare and Gross Indicator are locked at their current setting.
Graduations is forced to 1 Division and the menu item is hidden.
For EX…N…models, graduations will be locked at its current setting.
Weighing Units menu: all units are locked at their current settings.
Communication Menu:
Stable Weight Only is locked ON.
For EX…N…models, auto print mode selections are limited to OFF, On Stability,
and Interval. Continuous is not available.
Numeric Value Only is locked OFF.
Note:
The security switch must be in the locked position to set Legal for Trade to ON.
The security switch must be in the unlocked position to set Legal for Trade to OFF.
See Section 6.
5.6.9
Graduations
Set the displayed readability of the balance.
1 Division
= standard readability.
10 Divisions = readability is increased by a factor of 10.
For example, if the standard readability is 0.01g, selecting 10 Divisions will result in a
displayed reading of 0.1g.