EN-116
EXPLORER BALANCES
ON
= values are printed only when the stability criteria are met.
Print Numeric Value Only
Set the printing criteria.
OFF
= All data values are printed
ON
= Only numeric data values are printed
Print Options
Set the way to print.
Printer
= Print the result through a printer.
PC
= Transfer the result to a computer.
5.10.7 Auto Print
Set the automatic printing functionality.
OFF
= disabled
ON STABLE = printing occurs each time the stability criteria are met.
INTERVAL
= printing occurs at the defined time interval.
CONTINUOUS = printing occurs continuously.
When ON STABLE is selected, set the conditions for printing.
LOAD
= Prints when the displayed load is stable.
LOAD ZERO = Prints when the displayed load or zero reading is stable.
When INTERVAL is selected, set the time interval using the numeric keypad.
Settings of
1
to 3600 seconds are available.
5.10.8 Print Calibration Data
Set the automatic Calibration Data printing functionality.
OFF
= disabled
ON
= enabled
5.10.9 Print Content
Enter this sub-menu to define the content of the printed data.
See details below:
Header
Set the status.
OFF
= the user defined header is not printed.
ON
= the user defined header is printed.
USER DEFINED TEXT
USER DEFINED TEXT
USER DEFINED TEXT
USER DEFINED TEXT
USER DEFINED TEXT
Note:
The header information must be defined
using the H x “text” interface command.
(See Section 5.1.)
Date and Time
Set the status.
OFF
= the time and date are not printed.
ON
= the time and date are printed.
01/31/10 12:30 PM
Application Name
Set the status.
OFF
= the Application Name is not printed.
ON
= the Application Name is printed.
Sample Name
Set the sample name.
OFF
= the Sample Name is not printed.
ON
= the Sample Name is printed.
Batch ID
Set the batch ID
OFF
= the Batch ID is not printed.
ON
= the Batch ID is printed.
Application Name: XXXXXXXXXXXX
Sample Name: XXXXXXXXXXXX
Batch ID: XXXXXXXXXXXX