EXPLORER BALANCES
EN-123
5.12 Diagnostics
Use this menu to enable the Diagnostics (troubleshooting) functions, and to enter the Service menu;
5.12.1 Routine Test
To set settings for daily repeatability test and perform it.
You can set
Number of Repeatability
Tests
,
Test Weight ID
, T
est Weight
Value
and
Test Weight Class
in this
menu.
To perform repeatability test, press
Repeatability Test
and then follow
the onscreen instruction.
When the repeatability test finishes, it will pop out the results automatically. Press
to print
repeatability test report. Press
Min-weight
to show the reference minimum weight. After exiting
the routine test, the results will be deleted automatically.
5.12.2 Level Bubble Light
(not available in EX12001, EX24001, EX35001)
To verify proper operation of the illuminated Level Bubble.
When selected, this light should blink.
5.12.3 Draft Shield Light
(Analytical models)
To verify proper operation of the Draft Shield Light.
When selected, this light should blink.
5.12.4 Sensors
To verify proper operation of each Touchless Sensor.
When operated, each sensor should light and make a sound.
5.12.5 Automatic draftshield door
To verify proper operation of each Touchless Sensor.
When operated, each sensor should light and make a sound.
Note: Only for Automatic draftshield door models.