EXPLORER BALANCES
EN-151
9.4
Communication
9.4.1
Interface Commands
Commands listed in the following table will be acknowledged by the balance.
The balance will return “ES” for invalid commands.
TABLE 9-10. EXPLORER INTERFACE COMMAND LIST
Command
Characters
1)
Function
IP
Immediate Print of displayed weight (stable or unstable). Attention: for EX…N…models, when
LFT ON, IP could only print stable displayed weight.
P
1)
Print displayed weight according to “Stable only” setting in the communication menu. Attention:
when LFT ON, P could only print stable displayed weight.
CP
Continuous Print. For EX…N…models , when LFT ON, CP could not work.
SP
1)
Print on Stability.
SLP
Auto Print stable non-zero displayed weight. Attention: the corresponding settings in the
communication menu are modified too.
SLZP
Auto Print stable non-zero weight and stable zero reading. Attention: the corresponding
settings in the communication menu are modified too.
xP
Interval Print x = Print Interval (1-3600 sec) 0P ends interval Print. Attention: the corresponding
settings in the communication menu are modified too.
0P
See above.
H
Enter or get Print Header Lines. Attention: header text length is up to 25 characters, the setting
format is “H x “header line text””; and “H x” could get the corresponding header line in the
balance.
Z
Same as pressing Zero Key
T
Same as pressing Tare Key.
xT
Establish a preset Tare value in displayed unit. X = preset tare value. Sending 0T clears tare (if
allowed).
PT
Prints Tare weight stored in memory.
PM
Print current application mode (weighing mode).
xM
Set current application mode to x. x depends on application, use application list
2)
.
M
Scroll to the next enabled mode.
PU
Print Current weighing unit: g, Kg, lb, oz, etc….
xU
Set balance to unit x: g, Kg etc. x depends on unit list
3)
.
U
Scroll to the next enabled unit.
ON
Brings out of Standby
OFF
Goes to Standby.
C
Begin Span Calibration, same as trigger from calibration menu. Attention: when LFT ON, the
operation is not allowed.
IC
Begin internal Calibration, same as trigger from calibration menu.
UC
User Calibration (uses default weight), same as trigger from calibration menu. Attention: when
LFT ON, the operation is not allowed.
AC
Abort Calibration. Attention: when LFT ON, the operation is not allowed.
xUC
Set user defined weight and trigger one user calibration. Attention: the user defined weight is
only used for this command.
WI 1 0
Left door open, right door closed.
WI 0 1
Left door closed, right door open.
WI 1 1
Both doors open.
WI 0 0
Both doors closed.