EXPLORER BALANCES
EN-155
Balance Setup for USB Interface
A) The Ohaus Interface is preset to communicate using the following settings:
9600 baud, 8 bit, no parity, no handshake, 1 stop bit.
This is a fixed setting and cannot be changed.
B) Configure the balance to the desired USB and printing parameters.
SETUP DEFINITIONS
PRINT / Stable – On
Balance transmits stable data only.
PRINT / Auto Print – Continuous
Balance repeatedly sends data as fast as possible.
PRINT / Auto Print – On stability
Balance automatically outputs data upon stability.
PRINT / Auto Print – Interval
Balance sends data every (xx) seconds.
PRINT / Auto Print – Off
Only sends data when PRINT is pressed.
USB / Baud, Parity, Handshake
Set parameters to match printer or computer.
Operation
USB OUTPUT
With the Adapter installed the balance will operate in one of the following three ways:
USB = On, Auto Print = Off, Stable = On or Off
Press PRINT to send the display data based on the settings in the Menu.
If Stable is set to On, the balance will wait for a stable reading before sending data.
If balance has connect USB flash diver, the data will automatic go to USB.
USB = On, Auto Print = On, Stable = On or Off
The balance will automatically send data.
If Stable is set to On, only stable values will be sent.
USB = Off