CHAPTER 3 MAINTENANCE PROCEDURES
Ohaus Corporation www.ohaus.com
3-6
Ranger
™
(Count) 3000 Series Service Manual
TABLE 3-2. Ranger 3000/Ranger Count 3000 OVERLOAD STOP GAP SETTINGS
Model
Max. Capacity
Overload Stop Gap
R3000 / RC 3000 Scale
Load Cell
A/B mm
C/D mm
R31P(E)1502
RC31P1502
1.5kg
3kg
0.6
0.5
R31P(E)3
RC31P3
3kg
5kg
0.8
0.7
R31P(E)6
RC31P6
6kg
10kg
1.5
1.1
R31P(E)15
RC31P15
15kg
20kg
2.4
1.7
R31P(E)30
RC31P30
30kg
40kg
2.4
1.7
Figure 3-6. Gap position of corners A, B, C, D. (A & B represent the front of the scale.)
A
B
D
C
Load Cell