CHAPTER 4 TESTING
Ranger
™
(Count) 3000 Series Service Manual 4-1
Ohaus Corporation www.ohaus.com
4.1 TESTING
Before and after servicing a Ranger 3000/Ranger Count 3000 scale, an operational test and
various performance tests should be made to confirm that the scale meets specifications. Turn
the scale on and allow it to warm up for at least one hour before performing these tests.
NOTE:
Make sure the test area is free from drafts and that the scale rests on a level and
vibration-free surface.
4.1.1 TEST MASSES REQUIRED
The masses required to test the Ohaus Ranger 3000/Ranger Count 3000 scales must meet the
requirements of ASTM Class 4 or OIML F2 Tolerance. The mass values are listed in Table 4-1.
TABLE 4-1. TEST MASS VALUES
Model
Weight (g)
R31P(E)1502
RC31P1502
1kg,500g,200g
R31P(E)3
RC31P3
1kg,500g
R31P(E)6
RC31P6
2kg,1kg,500g
R31P(E)15
RC31P15
10kg,5kg,2kg
R31P(E)30
RC31P30
10kg,5kg
4.2
Operational Test
1. Connect a functioning Power cord to the back of the scale.
2. Plug the Power Cord into a suitable power source, or power the scale on using
battery power. (Assure that the battery is charged beforehand.)
4.3
Segment Display Test
Turn the scale on, and ensure that all segments are enabled and displayed briefly and
followed by a software revision number.