EN-15
3000 Series Indicators
3.2 Menu Navigation
3.3 Calibration Menu
Two calibration processes are available: Span Calibration and Linearity
Calibration.
Span Perform
Linearity Perform
Geographic
Adjustment Set 00…
Set 19
… Set 31
End Calibration Exit CALIBRATE menu
NOTES
:
1. Make sure that appropriate calibration masses are available before
beginning calibration.
2. Make sure that the scale base is level and stable during the entire
calibration process.
3. Calibration is unavailable with LFT set to On.
4. Allow the Indicator to warm up for approximately 5 minutes after
stabilizing to room temperature.
5. To abort calibration, press the
Exit
button anytime during the calibration
process.
TO ENTER THE MENU MODE
Press and hold the Menu button until MENU appears on the display. The first upper level menu appears on the display.
Summary of button navigation functions in menu mode:
--
Yes
Allows entry into the displayed menu.
- Accepts the displayed setting and advances to the next menu item.
--
No
Skips by the displayed menu.
- Rejects the displayed setting or menu item and advances to the next available item.
--
Back
Moves backwards through the upper and middle level menus.
- Backs out of a list of selectable items to the previous middle level menu.
--
Exit
Exits from menu directly to the active weighing mode.
Summary of Contents for Ranger 3000 Series
Page 1: ...i Indicatori serie 3000 Manuale di istruzioni Indicatore T31XW Indicatore T31P...
Page 2: ...ii...
Page 45: ...i Manuel d instructions pour les indicateurs de la s rie 3000 Indicateur T31XW Indicateur T31P...
Page 46: ...ii...
Page 89: ...i Indicadores serie 3000 Manual de instrucciones Indicador T31XW Indicador T31P...
Page 90: ...ii...
Page 133: ...3000 Series Indicators Instruction Manual T31P Indicator T31XW Indicator...
Page 134: ......
Page 179: ...i Bedienungsanleitung f r Indikatoren der Serie 3000 T31XW Indikator T31P Indikator...
Page 180: ...ii...