EN-25
3000 Series Indicators
3.7.7 End Unit
Advance to the next menu.
Reset
No
, Yes
Baud Rate: 300, 600, 1200, 2400, 4800,
9600
, 19200
Parity: 7 Even, 7 Odd, 7 None,
8 None
Stop Bit
1
or 2
Handshake: Off,
XON/XOFF
Stable Only
Off
, On
Auto Print
Off
,
On Stable (-> Load, Load and Zero),
Interval (-> 1…3600), Continuous
Content Gross (->
Off
, On)
Net (->
Off
, On)
Tare (->
Off
, On)
Unit (->
Off
, On)
End Print
Exit PRINT menu
3.8.1 Reset
Set the Print menu to factory defaults.
NO
= not reset.
YES
= reset.
3.8 Print Menu
Enter this menu to define printing parameters. Default settings are bold.
NOTE
: If the Legal for Trade menu item is set to ON, the following
settings are not reset: Stable, Auto Print
3.8.2 Baud
Set the Baud rate.
300 = 300 bps
600 = 600 bps
1200 =1200 bps
2400 = 2400 bps
4800 = 4800 bps
9600 = 9600 bps
19200
= 19200 bps
3.8.3 Parity
Set the data bits and parity.
7 EVEN = 7 data bits, even parity.
7 Odd = 7 data bits, odd parity.
7 NONE = 7 data bits, no parity.
8 NONE = 8 data bits, no parity.
Summary of Contents for Ranger 3000 Series
Page 1: ...i Indicatori serie 3000 Manuale di istruzioni Indicatore T31XW Indicatore T31P...
Page 2: ...ii...
Page 45: ...i Manuel d instructions pour les indicateurs de la s rie 3000 Indicateur T31XW Indicateur T31P...
Page 46: ...ii...
Page 89: ...i Indicadores serie 3000 Manual de instrucciones Indicador T31XW Indicador T31P...
Page 90: ...ii...
Page 133: ...3000 Series Indicators Instruction Manual T31P Indicator T31XW Indicator...
Page 134: ......
Page 179: ...i Bedienungsanleitung f r Indikatoren der Serie 3000 T31XW Indikator T31P Indikator...
Page 180: ...ii...