EN-33
3000 Series Indicators
6. LEGAL FOR TRADE
When the indicator is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with
local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are
met.
6.1 Settings
Before verification and sealing, perform the following steps:
1. Verify that the menu settings meet the local weights and measures regulations.
2. Perform a calibration.
3. Set Legal for Trade to ON in the Setup menu.
4. Exit the menu.
5. Disconnect power from the indicator and open the housing as explained in Section 2.3.1.
6. Set the position of the security switch to ON as shown in Section 1.2, Figure 1-3, Item 9.
7. Close the housing.
8. Reconnect power and turn the indicator on.
NOTE:
For installations that employ the audit trail sealing method, steps 5 to 8 are not required. However, the security switch
may be set to ON to safeguard against unintentional changes to configuration and calibration settings.
NOTE:
When Legal for Trade is set to ON and the security switch is set to ON, the following menu settings cannot be changed:
Span Calibration, Linearity Calibration, GEO, LFT, Calibration Unit, Capacity, Graduation, Power On Unit, Zero Range, Auto Zero
Tracking, Expanded Mode, Count Mode, Kilogram Unit, Pound Unit, Gram Unit, Ounce Unit, Pound Ounce Unit, Stable Only. To
enable editing of these menu settings, return the security switch to the off position and set LFT menu item to off.
6.2 Verification
The local weights and measures official or authorized service agent must perform the verification procedure. Please contact your
local weights and measures office for further details.
6.3 Sealing
6.3.1 Physical Seals
For jurisdictions that use the physical sealing method, the local weights and measures official or authorized service agent must
apply a security seal to prevent tampering with the settings. Refer to the illustrations below for sealing methods.
Figure 6-1. T31P Wire Seal Figure 6-2. T31P Paper Seal
Summary of Contents for Ranger 3000 Series
Page 1: ...i Indicatori serie 3000 Manuale di istruzioni Indicatore T31XW Indicatore T31P...
Page 2: ...ii...
Page 45: ...i Manuel d instructions pour les indicateurs de la s rie 3000 Indicateur T31XW Indicateur T31P...
Page 46: ...ii...
Page 89: ...i Indicadores serie 3000 Manual de instrucciones Indicador T31XW Indicador T31P...
Page 90: ...ii...
Page 133: ...3000 Series Indicators Instruction Manual T31P Indicator T31XW Indicator...
Page 134: ......
Page 179: ...i Bedienungsanleitung f r Indikatoren der Serie 3000 T31XW Indikator T31P Indikator...
Page 180: ...ii...