Ranger 7000
®
Series Service Manual
31
Ohaus Corporation www.ohaus.com
6. After removing the 3 cables mention above the next step would be removing the 4 screws which
are securing the MFR load cell from the base bottom housing.
Once these 4 screws are removed you can securely remove the cell from the base.
Location of the screws securing the MFR cell from the base bottom housing.
Re-Assembly:
Reverse the disassembly procedure to assemble the scale. Ensure that there is no foreign material
between the Base mounting surfaces and the Load cell.
Take special care to route the wires and cables according to their original positions. Proper routing
is important for RFI/ESD performance. Ensure that the wires or cables are not pinched during re-
assembly.