3.7.4 Removing MFR LFT PCBA, R71. (R71MHD3, R71MHD6, R71MHD15 & R71MHD35)
Model R71MHD6 is shown below.
1. Remove the 5 cables from the LFT PCBA.
2. Loosen the 2 screws securing the LFT PCBA with the base bottom housing.
1. Remove the below 5 cables from the LFT PCBA.
Location for the 5 cables.
A Cable, Cable link from LFT PCBA to optional rechargeable battery.
B Cable, Cable link from LFT PCBA to ON OFF Switch.
C Cable, Cable link from LFT PCBA to Main PCBA.
D Cable, Black/While cable link from LFT PCBA to AC Adapter.
E Cable, Red/White Cable from LFT PCBA to Main PCBA.
Summary of Contents for Ranger 7000 R71MHD3