B.4
COM Port Configuration.
1. Configure the software Com port configuration to match with the scale and click 'Next'.
Figure B-5. COM Port Configuration.
Note: Make sure there is no lose connection from R71 RS232 Port during the communication process.
2. Enter the scale’s Serial Number and IDNR number.
3. Enter the path to the image file, or click the Browse button to locate the file.
4. Click the Next button.
5. The software will indicate the download progress. When complete, disconnect the power from
the back of the scale, then re-connect.
6. Perform Service Calibrations (see Chapter 6). Test the scale.