Ranger™ Count 3000 Series EN-11
4. MENU SETTINGS
The User Menu allows the customizing of scale settings.
Note: Additional Sub-Menus may be available if Interface Options are installed. See Interface
User Manual for the additional setting information.
4.1 Menu Navigation
User Menu:
Menu:
C.A.L
S.E.t.U.P
r.E.A.d
U.n.i.t
Menu
Items:
Zero2
1
Span1
Span2
1
Lin1
Lin2
1
GEO
End
Reset
D.Scale
Cap2
1
Grad2
1
Pwr. Un
Zero
A.Opt
A.Tare
Bp.Opt
Bp.Sig
Bp.Key
Accum
End
Reset
Stable
Filter
AZT
Light
A.Off
End
Reset
kg
g
lb
oz
lb:oz
End
P.r.i.n.t.1
C.O.M.1
L.O.C.k
E.n.d
Menu
Items:
Reset
Stable
A.Print
Contnt
Layout
Data.Tr
End
Reset
Baud
Parity
Stop
Handsh
Alt.Cm
End
L.Cal
L.Setup
L.Read
L.Unit
L.Print
L.COM
End
Notes:
1
Only visible when dual scale is set to On.
Some units may not be available in all models.
When LEGAL FOR TRADE is set to ON (LFT Switch in locked position), the
menu settings are affected as follows:
Calibration (
C.A.L
) menu is not accessible.
Dual Scale, Capacity 2, Graduation 2, Filter and Units are locked at their current settings.
Zero Range setting is locked at 2%.
Stable Range setting is locked at 1d.
Auto-Zero Tracking setting is locked at 0.5d.
Stable Only is locked On.
Auto Print/ Continuous is disabled
Lb:oz is locked Off.