Ranger™ Count 3000 Series                                                          EN-1

 

1. INTRODUCTION 

This manual contains installation, operation and maintenance instructions for the Ranger™ 

Count 3000 Series. Please read the manual completely before using the scale. 

 

1.1 Definition of Signal Warnings and Symbols 

Safety notes are marked with signal words and warning symbols. These show safety issues and 

warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, 

malfunctions and false results. 

Signal Words   

CAUTION 

for a hazardous situation with low risk, resulting in damage to the device or 

  

 

the property or in loss of data, or injuries if not avoided. 

 

Note 

 

(No symbol) 

  

 

For useful information about the product 

Warning Symbols   

   

  

 

Attention Symbol 

 

 

               

Alternating Current 

 
1.2 Safety Precautions 

Please follow these safety precautions: 

  Verify that the AC input voltage printed on the data label matches the local AC power 

supply.  

  Do not drop loads on the platform.  

  Make sure that the power cord does not pose a potential obstacle or tripping hazard. 

  Do not position the scale in a way that makes it difficult to pull the power plug. 

  Use only approved accessories and peripherals. 

  Operate the scale only under ambient conditions specified in these instructions. 

  Disconnect the scale from the power supply when cleaning. 

  Do not operate the scale in hazardous or unstable environments. 

  Do not immerse the scale in water or other liquids. 

  Do not place the scale upside down on the platform. 

  Only use weights within the scale’s capacity as specified in these instructions. 

 

Service should be performed only by authorized personnel.

 

 

 

2. INSTALLATION 

2.1 Package Contents 

 

 Scale 

 Pan 

 Power Cord 

 Sub-platform 

 Instruction Manual / CD 

 Warranty Card 

 

 
 
 
 
 

Summary of Contents for Ranger Count RC1P1502

Page 1: ...Ranger Count 3000 Series Instruction Manual...

Page 2: ......

Page 3: ...ty Precautions Please follow these safety precautions Verify that the AC input voltage printed on the data label matches the local AC power supply Do not drop loads on the platform Make sure that the...

Page 4: ...firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Allow sufficient space around the scale 2 4 Leveling the Equipment The Ranger Count...

Page 5: ...eration the battery symbol indicates the battery charge level The scale will automatically turn off when the batteries are fully discharged For maximum operating time the battery should be charged at...

Page 6: ...s2 Off Turn scale off Units Change the weighing unit Switch between Check Weigh Check Count and Off Menu Enter user menu View the Audit Trail event counters extended press Menu Function Short Press Ye...

Page 7: ...Zero symbol 11 Auto Optimization Symbol 3 Negative symbol 12 Dynamic symbol not used 4 Stable weight symbol 13 Battery charge symbol 5 Preset Tare Tare symbols 14 Pound Ounce Pound Ounce symbols 6 Poi...

Page 8: ...ed in Checkweigh mode section 3 5 and will light up according to the following rules Red Loads Upper limit Green Loads Lower limit and Upper limit Yellow Loads Lower limit Figure 3 3 Below view of Ran...

Page 9: ...0 Positive Sampling 1 Place the sample on the pan 2 Input the sample size with the numeric keypad The value is displayed in the count window Alternatively the Sample key can be used instead Pressing...

Page 10: ...o display the statistical information To clear the accumulation data press the Tare key while the statistical information is displayed The display shows Clr aCC Press the Yes key to clear the stored d...

Page 11: ...under the target pieces range the yellow LED will light If the sample is within the target pieces range the green LED will light If the sample is over the target pieces range the red LED will light P...

Page 12: ...y reCall and lib are displayed in the weight and piece weight windows respectively 2 Long press the ID key edit and lib are momentarily displayed in the weight and piece weight windows respectively 3...

Page 13: ...nd P r i n t 1 C O M 1 L O C k E n d Menu Items Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Unit L Print L COM End Notes 1 Only visib...

Page 14: ...setting and advances to the next item No Skips by the displayed menu Rejects the displayed setting or menu item and advances to the next available item Back Moves backwards through the upper and middl...

Page 15: ...the span calibration 0 kg will be displayed Press Yes to accept C will be displayed while zero reading is stored Next the display shows the calibration weight value Place the specified calibration ma...

Page 16: ...tiates a 2nd scale linearity calibration procedure GEO GEO Geographical Adjustment Factor GEO is used to adjust the calibration based on the current location Settings from 0 to 31 are available with 1...

Page 17: ...layed at startup AUTO last unit in use when turned off kg kilograms g grams lb pounds oz ounces lb oz pound ounces Zero Range ZErO Set the percentage of scale capacity that may be zeroed 2 zero up to...

Page 18: ...is disabled ACCEPT the beeper will sound when the weight is within the Accept range UNDER the beeper will sound when the weight is below the Under setting OVER the beeper will sound when the weight is...

Page 19: ...ill maintain zero until a change of 0 5 divisions per second has been exceeded 1d the display will maintain zero until a change of 1 divisions per second has been exceeded 3d the display will maintain...

Page 20: ...t 3000 Series 4 5 Unit Menu This sub menu activates units so they will be accessible with the Units button The units in the menu must be turned on to be active Note Available units vary by model and l...

Page 21: ...printing occurs at the defined interval CONT printing occurs continuously ACCEPT printing occurs each time the display is within the Checkweigh accept range and stability criteria are met Content CON...

Page 22: ...is printed Header Set the status OFF disabled ON the Header is printed Note See section 7 3 1 for how to enter Header line Footer Set the status OFF disabled ON the Footer is printed Note See section...

Page 23: ...paper up four lines after printing FORM a form feed is appended to the printout Data Transfer data tr Output weighing results directly to a PC application OFF disabled ON enabled Notes Data Transfer F...

Page 24: ...ps 19200 19200 bps Parity paritY Set the data bits and parity 7 EVEN 7 data bits even parity 7 ODD 7 data bits odd parity 7 NONE 7 data bits no parity 8 NONE 8 data bits no parity Stop bit stOP Set th...

Page 25: ...ch can lock Menu settings to prevent tampering Default settings are in bold Reset reset Reset the Lock menu to factory defaults NO no reset YES reset Lock Cal l CAL Set the status OFF Calibration menu...

Page 26: ...n to the current application mode 4 10 Additional Features To use the weigh below feature remove AC power from the scale and remove the protective cover for the weigh below opening Install a suitable...

Page 27: ...2 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13 12 44 32 45 38 18 17 17 16 16 15 15 14 14 13 13 45 38 46 45 18 18 17 17 16 16 15 15 14 14 13 46 45 47 51 19...

Page 28: ...lained in Section 4 2 3 Set the switch to Locked See figure 5 2 The Menu Lock switch limits changes to the Cal Setup Readout Unit and Print menus The switch in type approved models may set some scale...

Page 29: ...he following menu settings are changed Dual Scale Capacity 2 Graduation 2 Zero Range Stable Range Auto Zero Tracking AZT Units kg g oz lb or lb oz or Stable Printing Dual Scale Capacity 2 Graduation 2...

Page 30: ...Err 8 1 Power On Error Weight reading exceeds Power On Zero limit Err 8 2 Power On Error Weight reading below Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4...

Page 31: ...526 0659 between 8 00 AM and 5 00 PM EST An OHAUS product service specialist will be available to provide assistance Outside the USA please visit our web site www ohaus com to locate the OHAUS office...

Page 32: ...20g 0 04lb Minimum Recommended APW 0 005g 0 00001lb 0 01g 0 00002lb 0 02g 0 00005lb 0 05g 0 0001lb 0 1 g 0 0002lb Capacity x Readability Max x e approved 1 5 kg x 0 0005 kg 1500 g x 0 5 g 3 lb x 0 00...

Page 33: ...Ranger Count 3000 Series EN 31 7 2 Drawings Figure 7 1 Dimensions 311 mm 12 2 in 327 mm 12 9 in 117 mm 4 6 in...

Page 34: ...ame as pressing Tare Key xT Download Tare value in grams positive values only Sending 0T clears tare if allowed PU Print current unit g kg lb oz lb oz xU Set scale to unit x 1 g 2 kg 3 lb 4 oz 5 lb oz...

Page 35: ...eight string print format Check Weighing application Field Weight Right aligned Space Unit Right aligned Space Stability Space T N G PT Right aligned Space Application Status Right aligned Term Length...

Page 36: ...n M metrology sticker on the descriptive data plate They may be put into service immediately Weighing Instruments to be verified in two stages have no green M metrology on the descriptive data plate a...

Page 37: ...online at europe ohaus com europe en home support weee aspx Thank you for your contribution to environmental protection FCC Note This equipment has been tested and found to comply with the limits for...

Page 38: ...has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other th...

Page 39: ......