EN-10
Ranger™ 3000
Series
Zero Check
Zero check is used when comparing subsequent samples to an initial reference sample. In this
case, the UNDER limit must be a negative value and the OVER limit must be a positive value.
Place the reference item on the scale and press
TARE
. Remove the reference sample and
place the item to be compared on the scale to determine if it is within the ACCEPT range.
3.7 Dynamic Mode
This mode allows the user to weigh an unstable load such as a moving animal. The weight is
held on the display until reset. Manual, semi-automatic and automatic start/reset methods are
available.
Manual operation
(DYNAMIC is set to MANUAL in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready) will then
be displayed.
2. Place the load on the pan and press the
Function
key to start measurement.
During the average period, the countdown timer decreases in one second increments.
3. After completed countdown, the average weight will be displayed. The tilde symbol will
blink indicating that the current weight is being held.
4. Manually reset the countdown timer by pressing the
Function
key. [
rEAdY
] (Ready) will
then be displayed.
Note
: If Set0 has been selected, the countdown timer is not displayed. The first weight larger
than 5d will be displayed and held.
Semi-automatic operation
(DYNAMIC is set to SEMI in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready) will then
be displayed.
2. Place the load on the pan to start measurement.
During the average period, the countdown timer decreases in one second increments.
3. After completed countdown, the average weight will be displayed. The tilde symbol will
blink indicating that the current weight is being held.
4. Manually reset the countdown timer by pressing the
Function
key. [
rEAdY
] (Ready) will
then be displayed.
Note
: If Set0 has been selected, the countdown timer is not displayed. The first weight larger
than 5d will be displayed and held.
Automatic operation
(DYNAMIC is set to AUTOMATIC in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready) will then
be displayed.
2. Place the load on the pan to start measurement.
During the average period, the countdown timer decreases in one second increments.
3. After completed countdown, the average weight will be displayed. The tilde symbol will
blink indicating that the current weight is being held.
4. Remove the load and the display will reset after 10 seconds, [
rEAdY
] (Ready) will then be
displayed.
Notes
: If Set0 has been selected, the countdown timer is not displayed. The first weight larger
than 5d will be displayed and held.
Alternatively, the display may be manually reset by pressing the
Function
key.