EN-16
Ranger™ 3000
Series
Reset
[
reset
]
Reset the Read menu to factory defaults.
NO
= not reset
YES
= reset
Stable Range
[
Stable
]
Set the amount the reading can vary while the stability symbol remains on.
0.5d
= 0.5 scale division
1d
= 1 scale division
2d
= 2 scale division
5d
= 5 scale division
Filter
[
FILtEr
]
Set the amount of signal filtering.
LOW
= less stability, faster stabilization time
MED
= normal stability, stabilization time
HI
= greater stability, slower stabilization time
AZT
[
AZt
]
Set the automatic zero tracking functionality.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 divisions per second has
been exceeded.
1d
= the display will maintain zero until a change of 1 divisions per second has been
exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been
exceeded.
Light
[
LIGHt
]
LCD models:
Sets backlight functionality.
OFF
= always off
ON
= always on
AUTO
= turns on when a button is pressed or the displayed weight changes.
LED models:
Set the light intensity.
LO
= low light intensity
MED
= medium light intensity
HI
= high light intensity
Sleep
[
SLEEP
]
Set the display shut off time.
OFF
= disabled (display will stay on until power off)
ON
= display will turn off in one minute in the absence of key or weight action
Auto off
[
A.OFF
]
Set the automatic shut off functionality.
OFF
= disabled
1
= powers off after 1 minute of no activity
5
= powers off after 5 minute of no activity
10
= powers off after 10 minute of no activity