Ranger RD
Appendix
7.5
.4 Verification
Before this product can be used in a trade approved application, it must be inspected
in accordance with local weights and measures regulations. It is the responsibility
of the purchaser to ensure that all pertinent legal requirements are met. Please
contact your local weights and measures office for further details.
7.5
.5 Sealing Scale
Scales must be certified by an accredited organization, and a certification seal must be affixed before the instrument may be used as a
certified scale.
7.5
.6 Unlocking Scale
When scales are in LFT mode, access to the technician mode (with the keypad) is blocked. To access the technician mode, proceed as
follows:
Switch off the scale or terminal.
Remove the bottom plate of the scale by undoing the retaining screws (Torx T20).
Caution: In order to remove the bottom plate it is necessary to break the certification seal! Once this seal has been destroyed, the scale
must be recertified by an accredited organization, and a new certification seal must be affixed before the instrument may be used as a
certified scale again!
Hold down the service switch (pushbutton) on the analog PCB (indicated by an arrow in the drawings below), with power applied to the
scale. Keep the service switch pressed in until “SCALE” appears in the display.
51
Summary of Contents for Ranger RD
Page 1: ...NSTRUCTION ANUAL 2ANGER 2 OMPACT 3CALES...
Page 2: ......
Page 17: ...3ETTINGS IN THE MENU 2ANGER 2 VERVIEW EVEL EVEL EVEL EVEL EVEL EVEL 0AGE 3 D 30 9 5 T G kg g...
Page 18: ...INST PRN Options not Available at this time...
Page 19: ......
Page 23: ...INST PRN Manual Data Output Stable and Unstable to the printer with print...
Page 31: ...NTERFACE DESCRIPTION 5DQJHU 5...
Page 36: ...HWd h H 4ECHNICAL DATA AND ACCESSORIES...
Page 44: ...E L7BK I E CB 9bWii _ j b c e hWp _YWb bWj_jkZ e lWbk HWd h H PPENDIX...
Page 48: ...48 Ranger RD Appendix...