CALIBRATION
When the balance is operated for the first time, a calibration is recommended to
ensure accurate weighing results.
External Span Calibration
With balance turned on:
Step-1
Press and hold
Menu
until [
mMeNU
] (Menu) is displayed.
Step-2
Release the button, the display will show [
C.A.L
].
Step-3
Press
Yes
to accept. [
SpaN
] will then be shown.
Step-4
Press
Yes
to begin the span calibration. [
--C--
] will blink and the
Zero reading is stored automatically.
Step-5
The specified mass blinks on the display.
Place the specified span calibration mass on the pan or press
No
to select an alternate weight.
[
--C--
] will blink while the reading is stored.
The display will show [
done
] if the calibration was successful. The
balance returns to the previous application mode and is ready for
use.
Internal Calibration (not available for SJX/E models)
With balance turned on:
Step-1
Press and hold
Menu
until [
mMeNU
] (Menu) is displayed.
Step-2
Release the button, the display will show [
C.A.L
].
Step-3
Press
Yes
to accept. [
In.CAL
] will then be shown.
Step-4
Press
Yes
to begin the Internal calibration. Zero reading is
stored automatically.
Step-5
Pull the InCal handle while the
display show [
PULL
].
Step-6
Push the InCal handle while the
display show [
PUSH
].
Step-7
Repeat the step-5, and step-6 one
more time.
Step-8
The display will show [
done
] if the calibration was successful.
The balance returns to the previous application mode and is
ready for use.