EN-8 SCOUT
TM
SERIES BALANCES - SPX
3.8 Totalization Mode
This mode allows the user to store a series of weight measurements.
Totalize mode has been
initiated when the symbol
“
∑
” is displayed and the current unit is displayed.
Notes:
Only positive numbers are totalized.
1. Press and hold
Mode
until [
total
] (Totalization) is displayed. [
CLr.tot
] will then
be displayed.
2. Press
Yes
or
No
key to clear the current totalized data or not. When a weight is added to
the scale the value is displayed.
3. If required, place an empty container on the pan and press
Tare
.
Add the first item, its
weight is displayed. Press
Function
to store the weight, the “
∑
” symbol will flash and
the display will show the total weight.
4. Press
Tare
(or remove the weight in previous operation) and add the next item. The
scale will display its weight. Press
Function
to store its weight. The
“
∑
” symbol will
flash and the new total weight will be displayed.
5. Repeat step 4 for all of the items to be accumulated.
6. To clear the stored total press and hold
Mode
until [
total
] is displayed. When [
Clr. tot
]
is displayed, press
Yes
.
3.9 Hold Mode
There are two modes for the display hold:
- Peak Hold: allows the user to capture and store the highest stable weight value (>=5d).
- Display Hold (default): allows the user to capture and store the first stable weight value (>=5d).
Start
If no weight value is held on the display, press
Function
key to begin. The [
rEAdY
] (Ready) will
be displayed until a weight is added on the pan.
When the stable value is being held on the display, the Hold icon (
H
) will blink and the displayed
weight will not change.
Reset
If the pan is empty and a weight value was held on the display, a single short press of the
Function
key will clear the held value and show the new weight on the pan.
1. Press and hold
Mode
until
[
HoLd
] is displayed.
2. If required, place an empty container on the pan and press
Tare
. Zero value will then
display.
3. Press
Function
key to begin. The [
rEAdY
] (Ready) will be displayed.
4. Place samples to be weighed on the pan.
5. The stable value will be held on the display, the Hold icon (
H
) will blink.