SCOUT
TM
SERIES BALANCES - SPX EN-9
4. MENU SETTINGS
The User Menu allows the customizing of balance settings.
Note: Additional Sub-Menus may be available if Interface Options are installed. See Interface
User Manual for the additional setting information.
4.1 Menu Navigation
User Menu:
Sub-Menus
C.a.l
S.e.t.u.p
M.o.d.e
U.n.i.t
E.n.d
User Menu
Items:
Span
Lin
Reset
Yes/No
Filter
Low/Med/High
AZT
0.5/1/3/Off
StableRange
0.5/1/2//5
Backlight
Off/On/Auto
Auto Tare
Off/On/On-acc
Auto Off
Off/1/5/10
Reset
Yes/No
Weighing
On/Off
Counting
On/Off
Percent
On/Off
Check
On/Off
Totalize
On/Off
Hold
Disp/Peak/Off
Reset
g
kg
ct
N
oz
ozt
dwt
lb
lb:oz
grn
hkt*
sgt*
twt*
t*
tola/tical
End
End
End
End
Notes:
* Only available in certain regions.
When the Security Switch is in locked position, the Calibration function is hidden, Units are
locked to the current setting.
Security switch:
4.2 Calibration Menu
Enter this menu to perform calibrations.
Span
[
SpaN
]
Initiates a span calibration procedure (zero and span)
Lin
[
LIN
]
Initiates a linearity calibration procedure (zero, mid-point and span).
End Cal
[
End
]
Advance to the next menu or return to the top of the current menu.
Span:
Perform
Linearity:
Perform
End Calibration:
Exit menu
.
No/Back
Long press -
Enter menu
Short press -
Accept
Short press
–
Go to
Next/Prev.
menu
Yes
Exit
Short press
–
Exit
menu