3-4
CHAPTER 3 TROUBLESHOOTING
3.3 ERROR CODES
Scout balances are equipped with software which will display an error condition when
it occurs. Table 3-2 Error Codes, describes the various error codes which can appear
on the display and specifies the probable reason and remedy.
TABLE 3-2. ERROR CODES
DISPLAY
REASON
Err 0 Invalid checksum data. The balance may need recalibration, particularly linearity
calibration. If the error persists after recalibration, return the balance to the
factory for servicing.
Err 1 Sample being weighed exceeds the capacity of the balance. If an error occurs
when the sample is within the balance's capacity, balance may be incorrectly
calibrated. An underload such as the pan off of the balance could also display
Err1. Recalibrate the balance.
Err 4 Incorrect or no calibration mass used when performing calibration procedure.
Error will flash momentarily, then the balance will use previous calibration data.
Recalibrate correctly.
Err 5 Value is too large to display; exceeds 99999 counts. Possibly caused by an
invalid parts counting sample.
Err 6 Parts counting error - sample weighs <1d. Balance shows error then exits parts
counting.
Err 9 Internal data error. Return the balance to the factory for servicing.
Summary of Contents for SCOUT
Page 1: ...OhausCorporation 29HanoverRoad FlorhamPark NJ 07932 0900 ELECTRONIC BALANCES SERVICE MANUAL TM...
Page 6: ...i v...
Page 10: ...1 4 CHAPTER 1 INTRODUCTION...
Page 38: ...5 2 CHAPTER 5 DRAWINGS AND PARTS LISTS...
Page 39: ...5 3 CHAPTER 5 DRAWINGS AND PARTS LISTS Figure 5 1 Exploded View of Balance 5 3 5 4...
Page 43: ...SCOUT ELECTRONIC BALANCES P N 300040 010 R1198...
Page 45: ...e e e...
Page 47: ......