Ohaus Corporation
Model:.............................................................Scout
Service Training
Description:.......................................Common Faults
FLT_SC
6.14.00
Page 1 of 1
Error 1
Check the downstop screw
Deadload out of range - Perform a linearity cal. If this fails replace the transducer
Error 4
Check the downstop screw
Wrong cal weight used - Perform a linearity cal. If this fails replace the transducer
No Display
Check operation of ON switch. Short to test. If this fails, check U6.
Defective U6, remove from circuit and test (balance will display Err 9). If this fails, replace PCB.