CHAPTER 1
GETTING STARTED
1.6 CALIBRATION
For best results, calibrate the scale at regular intervals. Correct calibration weights must be
available before starting calibration. (See Specification Table). Multiple weights may be used to
equal the required calibration weight. Calibration is unavailable in Dynamic Weighing Mode.
CAUTION!
BE CAREFUL HANDLING CALIBRATION WEIGHTS: THEY ARE VERY
HEAVY. IMPROPER LIFTING METHODS OR MISUSE OF CALIBRATION
WEIGHTS MAY RESULT IN PERSONAL INJURY.
1. Allow time for the scale to stabilize after moving it from an area which is at a different
temperature than the area where it is to be operated: one hour for each 5°F (2.7°C)
change. After temperature stabilization, allow an additional 20 minutes after turning the
scale on, for the scale electronics to stabilize.
2. To select the weighing unit to be used, press
UNITS
Cal
until the correct unit is
displayed, then press
Yes
.
Note
: Calibration can be stopped at any time: press
UNITS
Cal
or turn the scale off.
3. Press and hold
UNITS
Cal
until
CAL
is displayed.
4. With the platform empty, press
ON/ZERO
Off
to capture the scale
zero. The display will show
-C-
.
5. After the zero is captured, the required span calibration weight
value in the selected unit appears.
6. Place the calibration weight on the platform and press
ON/ZERO
Off
.
The display will again show
-C-
.
7. After span capture, the display will return to the normal weighing
mode.
Note
: The message
CAL E
will appear if an incorrect calibration weight was applied.
Repeat the procedure using the correct calibration weight.
Calibration Locking
: By use of an internal switch, the calibration function may be locked
(disabled), and unauthorized recalibration prevented.
1. With the scale powered off, remove the back housing section of the scale by unscrewing
the four screws at the bottom.
2. Slide the switch on the PCB (marked
SW3
) to the position marked
CAL LOCK
.
3. Reassemble the housing.
4. If required, place tamper evident seals over the screw holes or the housing edges.
If later recalibration is required, the housing will need to be reopened and the lock switch
returned to the original position.
Note
: When the lock switch is set to the
CAL LOCK
position, the power-on zero range is
reduced to 10% of full capacity.
SD Series Scales Service Manual
1-5
Ohaus Corporation www.ohaus.com
Summary of Contents for SD Series
Page 2: ......
Page 4: ......
Page 12: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 6 SD Series Scales Service Manual ...
Page 20: ...CHAPTER 2 TROUBLESHOOTING Ohaus Corporation www ohaus com 2 8 SD Series Scales Service Manual ...
Page 36: ......
Page 37: ... 80120061 P N 80120061 SERVICE MANUAL SD SERIES BASE INDICATOR ...