SCOUT
TM
SERIES BALANCES - SKX EN-5
Figure 3-2. Scout Display
TABLE 3-2. Display Symbols
Item Description
Item Description
1
Battery charge symbol
7
Pointer symbols
2
Stable weight symbol
8
Kilogram, grain symbols
3
Negative symbol
9
Pound, Ounce, Pound:Ounce symbols
4
Center of Zero symbol
10
g/mol symbol
5
NET symbol
11
G symbol
6
Preset Tare symbol
12
Newton, Bluetooth symbols
3.2 Turning Balance On/Off
To turn the balance on, press and hold the
On/Zero Off
button for 1 second. The balance
performs a display test, momentarily displays the software version, and then enters the active
weighing mode.
To turn the balance off, press and hold the
On/Zero Off
button until OFF is displayed.
3.3 Calibration Operation
When the balance is operated for the first time, a span calibration is recommended to ensure
accurate weighing results. Before performing the calibration, be sure to have the appropriate
calibration weights. Ensure that the Security switch is set to unlocked position.
Press and hold Menu until [
mMeNU
] (Menu) is displayed. When the button is released, the display
will show [
C.A.L
]. Press
Yes
to accept. [
SpaN
] will then be shown. Press Yes to begin the span
calibration.
[
--C--
] will be displayed while zero reading is stored. Next, the display shows the calibration
weight value (press
No
to toggle value). Place the specified calibration mass on the pan. [
--C--
]
will be displayed while the reading is stored. The display will show [
done
] if the calibration was
successful. The balance returns to the previous application mode and is ready for use.
3.4 Weighing
Mode
This mode is the factory default setting.
1. If needed, press and hold
Mode
until [
wWeIGH
] (Weigh) is displayed.
2. If required, place an empty container on the pan and press
Tare
.
3. Add sample to the pan or container. The display shows the weight of the sample.
2
3
4
1
5
6
8
9
10
11
12
7