4. SETTINGS
4.1 Menu Mode and Navigation
To enter the Menu Mode
Press and hold the
SET
button for 3 seconds.
[
SET-F
x] will be displayed where x =:
0= set upper and lower limit value.
1= User Settings Menu.
2= restores settings to factory default.
3= set tare capacity between 2/3Max. and full scale.
To navigate in the Menu Mode:
Press
¨
to toggle through available settings.
Press
ENTER
to accept the displayed setting and advance to the next step.
Press
SET
at any time to save settings and return to weighing.
4.2 Set Upper and Lower Limit Values
Enter the Menu Mode (Section 4.1). [
SET-F0
] will be displayed.
Press
ENTER
to accept the displayed setting and view the upper limit [000000].
The Over LED will illuminate.
Press
¨
,
ª
,
©
to edit the upper limit.
Press
ENTER
to confirm the setting and display the lower limit [000000].
The Under LED will illuminate.
Press
¨
,
ª
,
©
to edit the lower limit.
Press
ENTER
to confirm the setting and the display will show the alarm setting
[
rUn
X], where x=:
0 = check weigh mode is disabled.
1 = under / accept / over LEDs are active, the beeper will sound when the weight
is outside the accept range.
2 = under / accept / over LEDs are active, the beeper will sound when the weight
is within accept range.
3 = under / accept / over LEDs are active, the beeper is disabled.
Press
¨
to change the alarm setting.
Press
ENTER
to confirm setting and display the upper limit.
Press
SET
to store the values and return to weighing.