T21P Indicator EN-13
5. SCALE SETUP AND CALIBRATION
NOTES:
1.
Make sure that appropriate calibration kilogram masses are available
before beginning calibration.
2.
Make sure that the scale base is level and stable during the entire
calibration process.
3.
Allow the Indicator to warm up for approximately 5 minutes after stabilizing
to room temperature.
4.
To abort calibration, press and hold the
SET
button for 3 seconds anytime
during the calibration process. Or turn unit off.
Turn on the indicator. During startup display sequence press and hold
SET
until
[
FS__SP
] is displayed.
Press
ENTER
to display [
P___
x] (x=position of decimal point: 0, 1, 2, 3, and 4).
Press
ª
,
©
to select the desired setting.
Press
ENTER
to display [
D___
x] (x=divisions: 1, 2,5,10, 20, and 50).
Press
ª
,
©
to select the division.
Press
ENTER
to display Maximum Capacity Setting.
Press
¨
,
ª
,
©
to set the desired value.
The LED row below the display refers to the active segment.
Press
ENTER
to accept and [
noLoad
] is displayed.
Make sure pan is empty.
Press
ENTER
to accept and [
Load
] is displayed.
Press
ENTER
to display [xxxxxx] (xxxxx= Span Weight).
Press
¨
,
ª
,
©
to set a different value
Place calibration weight on the pan
Press
ENTER
to accept and store data.
[
CALEnd
] will be displayed for 2 seconds and the indicator will enter weighing
mode.