EN-
5000 Series Indicators
.5. Filter
Set the amount of sgnal flterng.
LOW = less stablty, faster stablzaton tme (<1 sec.)
MEd = normal stablty, stablzaton tme (<2 sec.)
HI = greater stablty, slower stablzaton tme (<3 sec.)
.5. Auto-Zero Tracking
Set the automatc zero trackng functonalty.
OFF = dsabled.
0.5 d = the dsplay wll mantan zero untl a change of 0.5 dvsons per second has been
exceeded.
1 d = the dsplay wll mantan zero untl a change of 1 dvson per second has been
exceeded.
3 d = the dsplay wll mantan zero untl a change of 3 dvsons per second has been
exceeded.
NOTE
: When the LFT menu tem s set to ON, the selectons are lmted to 0.5d, 1d and 3d. The settng s
locked when the hardware lock swtch s set to the ON poston.
.5.1 Reset
Set the Readout menu to factory default settngs.
NO = not reset
YES = reset
If the Legal for Trade menu tem s set to ON, the Stable Range, Averagng Level, Auto Zero Trackng, Auto Off
and Gross settngs are not reset.
.5. Stable Range
Set the amount the readng can vary before the stablty symbol turns off.
0.5d = 0.5 scale dvson
1d = 1 scale dvson
2d = 2 scale dvsons
3d = 3 scale dvsons
5d = 5 scale dvsons
NOTE
: When LFT s set to ON, the settng s forced to 1 d. The settng s locked when the hardware lock
swtch s set to the ON poston.