Traveler
™
Scales
EN-4
2.6 Initial Calibration
Power on the scale by pressing
On-Zero
. Allow at least one minute for the scale to stabilize
before calibrating. Press and hold the
Print/Cal
button until [
CAL
CAL
CAL
CAL
] is displayed. The display
flashes [
----CCCC----
] while zero reading is taken. The display flashes the required span mass. Put
the indicated calibration mass on the pan and press
Yes.
The display flashes [
----CCCC----
]. When
calibration is complete, [
Don
Don
Don
DonEEEE
] is displayed.
Note:
Calibration is also available in the Setup menu, see section 4.2.1.
3. OPERATION
Some weighing Units of measure may initially be unavailable. They can be enabled in the
Main menu Unit [
U.n.i.t
], see section 4.
3.1 Draft Shield
The draft shield was designed with an easy to remove center piece which can be used as a
high quality sample tray. Even with the center piece removed the draft shield will provide
increased protection from rough environments while maintaining weighing speed. To
remove the center piece, rotate counter-clockwise. See Figure 3-1. With the sample tray
firmly in place the unique shape of the draft shield also allows for the efficient stacking of
one scale on top of another.
Figure 3.1, Sample Tray Removal