19
CALIBRATION MENU
CALIBRATION MENU
CALIBRATION MENU
CALIBRATION MENU
CALIBRATION MENU
Precision Plus balances can be calibrated in two ways: Span calibration or Linearity
calibration. Span calibration resets the balance’s weighing range using two weight
values: zero and a weight value at or near the balance’s capacity. Linearity calibration
minimizes deviation between actual and displayed weights within the balance’s
weighing range. Three weight values are used: zero, a weight value within the
balances weighing range, and a weight value at or near the balance’s specified
capacity. The following table shows the sequence in which submenus appear on the
Calibration menu.
CALIBRATION MENU TABLE
CALIBRATION MENU TABLE
CALIBRATION MENU TABLE
CALIBRATION MENU TABLE
CALIBRATION MENU TABLE
Selects span calibration.
Selects linearity calibration.
Used to exit the Calibration menu.
Calibration Menu Protection
Calibration Menu Protection
Calibration Menu Protection
Calibration Menu Protection
Calibration Menu Protection
Calibration may be locked out to prevent
unauthorized personnel from changing
calibration. To lock out calibration menu,
refer to the section titled Menu Lock-Out
Protection.
NOTE
NOTE
NOTE
NOTE
NOTE: If calibration has been locked out,
you will not be able to access it.
Calibration Masses
Calibration Masses
Calibration Masses
Calibration Masses
Calibration Masses
Before beginning calibration, make sure
masses are on hand. If you begin calibra-
tion and realize masses are not available,
either turn the balance off, or go through
the procedure without masses. The bal-
ance will use previously stored calibration
data. Calibration should be performed as
necessary to ensure accurate weighing.
Masses required to perform the proce-
dures are listed in the adjacent table.
CALIBRATION WEIGHTS
CALIBRATION WEIGHTS
CALIBRATION WEIGHTS
CALIBRATION WEIGHTS
CALIBRATION WEIGHTS
LINEARITY
LINEARITY
LINEARITY
LINEARITY
LINEARITY
SPAN ONLY
SPAN ONLY
SPAN ONLY
SPAN ONLY
SPAN ONLY
MODEL
MODEL
MODEL
MODEL
MODEL
MASSES
MASSES
MASSES
MASSES
MASSES
MASS
MASS
MASS
MASS
MASS
TP200
100g, 200g
200g
TP400
200g, 400g
400g
TP400D
50g, 400g
400g
TP600
200g, 500g
500g
TP2K
1kg, 2kg
2kg
TP4K
2kg, 4kg
4kg
TP4KD
500g, 4kg
4kg
Masses must meet or exceed ASTM
Class 1 Tolerance. Calibration masses
are available as accessories.
Summary of Contents for TP200S
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 27: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......
Page 52: ......
Page 53: ......
Page 112: ...Ohaus Corporation 29 Hanover Road Florham Park NJ 07932 0900 he...