23
Trooper Scale
Cal in kg:
Span cal choices
Linearity cal (fixed)
TR3RS
1,
2
, 3kg
2 & 3 kg
TR6RS
2,
4
, 6kg
4 & 6 kg
TR15RS
5,
10
, 15kg
10 & 15 kg
TR30RS
10,
20
, 30kg
20 & 30 kg
Cal in lb:
Span cal choices
Linearity cal (fixed)
TR3RS
2,
4
, 6 lb
4 & 6 lb
TR6RS
5,
10
, 15 lb
10 & 15 lb
TR15RS
10,
20
, 30 lb
20 & 30 lb
TR30RS
20,
40
, 60 lb
40 & 60 lb
5.CALIBRATION AND SEALING (Cont.)
Procedure (Cont.)
TURNING SOFTWARE LOCK OFF
For span calibration, there are a number of choices for
calibration. The highlighted number is the default.
To change the span calibration value read the setup sec-
tion of the menu on page 10. In the Setup menu, change
the CP selection to the desired value.
When the desired selection has been chosen and the
masses are available, you are ready to begin the calibra-
tion routine.
Procedure
SPAN CALIBRATION
With the Scale ON, press and hold the
G/N/T/
MENU
button until MENU is displayed. When you release t h e
G/N/T/
MENU
button, CAL is displayed. If the display
shows SETuP, read the section on turning off the Calibra-
tion Lock Switch or check the settings for legal for trade.
Press
G/N/T/
MENU
button, SPAN Y is displayed.
Press
G/N/T/
MENU
button, -C- is displayed. The scale
MUST be stable during this period and is establishing a
zero point. After a few seconds, the requested weight
value is displayed.
Summary of Contents for TROOPER SCALE
Page 4: ......
Page 100: ...32 Balance Trooper...
Page 102: ...34 Balance Trooper...
Page 103: ...Leerseite...